Haemoglobins
POCT Test Name
Haemoglobin
|
Instruments Hemocue monitors. |
|
Location within CDHB
Burwood Hospital- PACU (OR); Special Care Unit; Orthopaediac OPD- (Hemocue).
Christchurch Women's Hospital- Fetal Maternal Medicine, (Beckman Coulter Act-8 haematology analyser).
CWH- Birthing Suite; Gynae Day Stay Unit; (Hemocue)
Christchurch Hospital-Theatre 9 (OR); Recovery; (Hemocue).
Tests measured
Haemoglobin. Hemocue monitors are small transportable meters that utilise a disposable microcuvette. A few microlitres of blood from a capillary collect are aspirated into the microcuvette which is wiped carefully and then inserted into the Hemocue meter. Result in g/L in 60 seconds.
Sample and volume requirements
Hemocue- 10 µL is the sample volume.
Consumables
Hemocue Microcuvettes.
Controls
Hemocue Control cuvette, calibrated to each monitor. Used before each patient test.
Weekly correlations by the POCT Coordinator.
Reference Range
All reference ranges can be found in the CHLabs test database. Click on the left hand menu link to Canterbury Health Laboratories home page, then click on "Tests".
Sources of Error
Hemocue:
Poor capillary collection technique
AcT8:
Competencies
AcT-8 competencies are available.
Hemocue competencies are available.
Contact the POCT Coordinator for details.