|
There is confusion
in regarding the terms "calibration" and "quality
control", or QC (or IQC- Internal QC). Many POCT
operators talk about calibrating their POCT analyser or process when they are
actually analysing QC samples.
The processes of calibration and
quality control are two sides of the same coin: they complement each other. The
calibration establishes an initial point of measurement or data point in a
reaction; the QC checks that the calibration is correct. Together, they
determine the reliability of the method, i.e., the accuracy
and precision of the method.
The processes described on this
page apply to all laboratory testing, not just POCT.
Operating principles
Modern laboratory and POCT
analysers work on a variety of operating principles- spectrophotometry,
electrochemistry, potentiometry, amperometry, chemo- and electro-luminescence,
reflectometry, etc.
Regardless of their operating
principles, all measuring processes must be calibrated before use. The
simplest explanation of calibration deals with traditional spectrophotometric
principles.
Types of calibrators
There are also different types of
calibrators- some POCT analysers, e.g., Glucocard meters, or Roche CoaguChek
meters use a factory-set chip, replaced in the analyser each time a new box of
strips is opened. Lot numbers of calibrator and test strips must be identical so
as to eliminate any error caused by variation between lot numbers.
Some analysers like the i-STAT
analyser use a cartridge with a built-in calibrating solution- after the
patient's blood is loaded into the cartridge, the cartridge is inserted into the
analyser. The analyser mechanism causes a small reservoir of calibrator fluid to
move across the electrodes in the cartridge. After the electrodes are
calibrated, the patient's blood sample is shifted into position and analysed.
On the other hand, bench top
blood gas analysers require a separate liquid or gas calibrator to be analysed
in a process analogous to that of a patient sample.
Liquid calibrators can be made
from whole blood, serum or plasma or aqueous solutions, depending on the assay.
If you are interested in more
information about any of the above topics, please contact the POCT
Coordinator.
Calibration
A calibration
determines the initial value(s) in a reaction. These values are equivalent to
the known concentrations in the calibrating solutions. A minimum of 2
calibrators is required to establish a calibration factor, which is often
referred to as "k".
Calibrations are required when
the k value is either unknown or has shifted in value from a previous
calibration.
There are many reasons why this might have happened. Here are some of them:
-
a new analyser
-
a new test
-
a slow change in analytical
conditions. Sometimes referred to as "drift". There can be
various reasons for drift. It is usually detected when the QC fails.
-
new hardware, e.g., a
replacement electrode
-
new reagents
-
new calibrator
-
out of range QC
-
any other change in
analytical conditions
We can either make calibrators in
the lab (very carefully!), or we can purchase them from a manufacturer, the
usual procedure these days. (See the sidebar for more information).
The exact values of concentration
of each calibrator- there may be more than one analyte you are calibrating, say
in a blood gas analyser- are keyed into the analyser and stored there. The lot
number and expiry date is usually also stored. All these values will not be altered until a change of calibrator with different values,
lot number and expiry date is
introduced.
The calibrator is then analysed
in a manner identical to that of a patient sample but the values derived from
each reaction are stored in the analyser in special files relating to
calibration.
The analyser then calculates the
k
or calibration factor for each test and stores the value in its memory.
How it works
Of course, this procedure can be
done manually on the bench with a sheet of graph paper and a calculator. It used
to be done this way in laboratories all the time. However, these days we have
analysers.
With most
biochemical analyses, e.g., glucose if the absorbance is plotted against concentration, you would get a straight
line on a graph, i.e., a linear reaction:

The angle or
slope of the line gives you the k value, which remains constant for the whole of
the line length.
k is calculated using Beer-Lambert's Law. (see the
sidebar). This is a fundamental physical law which states that:
Given a linear spectrophotometric
reaction (see above), that:
Absorbance
is always proportional to Concentration.
Once the calibration factor is
included in the calculation,
Absorbance
multiplied by k equals Concentration.
Therefore, when a QC sample of
known value, or a patient's sample with an unknown value is analysed, its energy
value or Absorbance in the reaction is multiplied by k to give the
Concentration.
Here is another way of looking at
it. The graph on the left details the calibration procedure. The graph on the
right details QC or patient sample analysis. Note the direction of the red
arrows on each graph:
| CALIBRATION |
Q C
or PATIENT |
|

|

|
| 1. |
Value (Concentration) of calibrator is known. |
|
1. |
The Concentration of the QC or patient sample is not
known. |
| 2. |
Reaction rate (Absorbance) of Calibrator is measured. |
|
2. |
Reaction rate (Absorbance) of QC or patient sample is
measured. |
| 3. |
The calibration factor, k, is calculated
k = Concentration divided by Absorbance |
|
3. |
The Concentration is calculated.
Concentration = Absorbance multiplied by k |
Notes:
-
For Beer-Lambert's Law to
hold true, i.e., if "absorbance is proportional to
concentration", then a number of conditions must be included in
the calculation- the choice of reagents, the light
measuring capabilities of the analyser, the length of the measuring
chamber or cuvette, the wavelength chosen to
measure the reaction and so on.
-
k is
represented by the slope of the line on the graph. So long as the reaction stays linear, i.e., the line stays straight, k
will have an identical value for any particular reaction.
-
When the reaction is not
linear, things really get tricky and rather interesting! But that's
another story... :-)
Quality Control (QC)
Once we have calibrated our analyser, we still need to know if it has been
calibrated correctly. Thus the QC results check on the calibration. They
should tie in together, each proving the correctness of the other.
Click on this link for a far more detailed discussion on Quality
Control.
|
NOTE: The QC analysis checks the operator's ability
to perform the analysis and whether the correct result will be forthcoming
on the patient sample.
What it does not check for is the quality of the patient's
sample. The quality of the sample, especially if a capillary
whole blood sample, is fundamental to the generation of the correct
result.
|
|
Points to remember:
-
Calibrations
must be performed if analysis conditions alter in any way,
e.g., a change of consumables, hardware or reagents or if the QC
fails.
-
No
analyser should ever be used without a current calibration.
-
No
analyser should ever be used without a recent correct QC
result.
-
QC
samples should be analysed regularly- at least once per day.
-
All changes in procedure, reagents, lot numbers, etc;
maintenance and troubleshooting problems; and QC results must be
documented.
-
All error codes,
incorrect QC or procedural lapses MUST BE DOCUMENTED. Use the
Error Log associated with each analyser- each Error Log is to
be found in its associated Operators Manual .
|
|
|
What calibrates the calibrator? |
|
A
calibrator's value is determined by the "Reference Method",
the recognised most accurate and precise methodology for the
test in question.
Reference Methods
utilise high tech equipment like mass spectrometers, atomic absorption
(AA) spectrophotometers
and Gas Chromatograph (GC) or High Performance Liquid Chromatographs (HPLC).
The purest substances are used to obtain the most accurate assays.
Calibrator solutions, supplied by manufacturers, are made from these purest of
substances. They are analysed
repeatedly on most common laboratory analysers. The results are
statistically analysed to give the Average or Mean Value for each
constituent on each analyser type and often for different test kits, whose
test principle (and therefore sensitivity) may vary. These
values are then keyed into the appropriate laboratory analyser and a
calibration is performed for each analyte.
You may ask why all lab tests aren't analysed directly by Reference
Method. Cost and efficiency come into it. Often processes using mass
specs, AA or HPLC are elaborate, expensive and time consuming. Even so,
some lab tests are measured using these methodologies- drug assays or vitamins. Generally though, methods
are chosen that are simple and easy to use, but which still give reliable
accurate results.
The routine methods used must give values close to the
reference method in order to be accurate. However, different
methodologies do have different characteristics. That is why it is
important that each lab establishes their own range of values for their
QC material.
It is also important for each laboratory to participate
in External Quality Assurance programs, so that they can compare their
results to other labs to ensure that all results are identical within an
acceptable degree.
A
bit more detail...
The
commonest type of laboratory analyser is called a Spectrophotometer. In
its simplest form, it is an analyser in which a beam of monochromatic (single
coloured) light, adjustable to any particular light wavelength, can be passed
through a coloured solution derived from the analyte in question. An
example is urea which gives a blue colour using the commonest method. This
solution can be either the calibrator or patient/QC samples, depending on
whether the test is being calibrated or samples are being analysed.
Usually complementary coloured light is used,
e.g., a red beam would be passed through a green solution.
A
percentage of the light, called the Absorbance, is assimilated by
the coloured solution. The Absorbance is directly proportional to the Concentration of the
substance in
solution.
The emerging light, the Transmittance, is
inversely proportional to the Concentration. It is
measured by a detector, converted to
a digital signal, processed electronically and the expression is finally multiplied by the
Calibration Factor. The result can then be
displayed on a screen or can be printed out.
(In most
calculations, even though the transmittance (t) is actually measured, the
Absorbance (a) has more relevance as this is directly related to the
Concentration (c).
a
1/t
c
1/t
a
c
Concentration
The value of the substance under investigation. Various units of
measurement are g/L, mmol/L, umol/L, IU/L, nmol/mL and so on.
|
Beer-Lambert's
Law
A°
C or
A°. k = C
where:
A° is Absorbance
is
the symbol for "proportional to"
C
is Concentration
.
is
the symbol for "multiplied by"
k
is konstant or calibration factor |
|
|