Urine Test Strips

Click on the links to navigate this page
Introduction Correct timing of  colour development
General information Correct colour interpretation
Correct specimen collection Correct sample disposal
Patient and operator identification Limitations of procedure 
Correct  technique Quality Control

Competency documents, both paper and electronic are available for those wishing to update their records. A Competency is also available for Clinitek 50 urine reflectance meters.

Introduction

The CDHB uses two types of urine dipstick, the Roche L Combur-7 and the Bayer Multistix 8. Most wards use L Combur-7 strips. Those using Bayer Clinitek 50 reflectance meters for their urine testing must use the Multistix 8 strips.

Combur-7 strips are available from the Christchurch Pharmacy.

See the chart below for the list of components of each type- (Y= tests available; N= not).

Component Roche L Combur-7 Bayer
Multi-Stix 8
pH Y Y
protein Y Y
glucose Y Y
ketones Y Y
blood (rbc) Y Y
nitrite Y Y
leucocytes (wbc) Y Y
specific gravity (SG) N Y
urobilinogen N N
bilirubin N N

 General information


Correct specimen collection


Patient and operator identification

 

Patients must be correctly identified with a minimum of 2 points of ID- preferably name and national patient number. Temporary admission ID and date of birth are also satisfactory. 

The staff member performing the measurement must also be identified. Ensure that all printouts or transcriptions are signed clearly and dated

 


Correct strip immersion time and technique

 

This is very important- too little or incomplete immersion will cause insufficient penetration of urine into the test strip. This will result in a possible false negative result.

Too long an immersion time will result in elution or washing out of the test strip reagent colour, resulting in a pale reaction or falsely low result.

Read the instructions carefully- The general rules are:

  1. Collect FRESH urine in a clean dry container.

  2. Remove one strip and immediately replace the lid tightly.

  3. Briefly (no longer than 1 second) dip the whole of the reaction area of the strip into the urine.

  4. Wipe the side edge of the test strip along the edge of the vessel to remove excess urine.

  5. After the appropriate time (usually 60 seconds, but time may vary for same components), in a good light such as fluorescent, compare the colour reaction with the colour charts on the label. 

  6. Lay the urine test strip alongside the colour charts. Avoid laying the strip directly on the colour charts as this will result in the urine soiling the chart.


Correct timing of each component colour development

 

As noted above, correct timing is very important. Too little time to read may result in false negative readings; too much delay in reading time may result in false positives.

 

  • For the Combur-7 strips, read all colours at 60 seconds.
  • For the Multistix 8, read the glucose at 30 seconds; the specific gravity at 45 seconds; the pH, protein, blood and nitrite at 60 seconds and the leucocytes at 2 minutes.
  • Any colour changes that occur after 2 minutes are of no consequence.

 


Correct colour interpretation

 

Always compare the reagent strip colours side by side with the colour scales on the label, in a good light, e.g., fluorescent light. Do not read the strips in direct sunlight.

If the strips are read on an instrument, e.g., a urine test strip reflectance meter, always follow the instructions in the appropriate operator's manual.

(An interesting note: In the last 12 months or so during their laboratory inspections, the American accreditation authorities, (CLIA, JCAHO and CAP) have focused on colour blindness testing of all POCT operators throughout the US.) 

 


Correct sample disposal

 

Always dispose of samples and contaminated consumables in a recommended waste disposal container. This should be in accordance with recommended CDHB Infection Control and Health and Safety guidelines.

 


Limitations of procedure 

 

Here is a very interesting journal article detailing errors resulting from human versus reflectance meter colour interpretation: "Error Rates" using different reading methods. (This is a Word document. Use the Back button on your browser to return here once you have finished reading the article).

 

Many of the problems associated with manual colour interpretation and timing of the readings can be overcome by the use of a test strip reader. The one in use in several locations within the CDHB is the Clinitek 50 reflectance meter.

 


Problems associated with each component

Glucose

Ketone

Specific Gravity (S.G.)

Blood

pH

Protein

Nitrite

White cells (leucocytes)


Quality Control

The performance of reagent test strips should be monitored regularly, say once per week, with known negative and positive samples or a quality control (QC) test.

An example is the Bayer CHEK-STIX® quality control solution. This is purchased by the lab as a vial of special QC dry strips. Each QC strip is soaked in 12 mL of distilled or deionised water for 30 minutes. The dyes from the QC strip leach out into the water. The QC strip is then discarded and the solution is mixed well. This gives a "urine" of known values. The solution is stable for 7 days at 4 degrees Celsius. It may also be aliquoted and frozen.

In fact, all Clinitek 50 users are supplied with frozen aliquots of made up solution. A fresh vial is thawed, mixed and analysed daily or as required. Results are monitored and written on the QC charts provided. 

This solution is analysed using a regular urine test strip in an identical manner to a patient's sample. The printout values should fall within predetermined areas on the CHEK-STIX control chart as supplied by the POCT Coordinator. The correct colour match is proof that the test strips are still working satisfactorily. Only if the QC results are correct, are patients' samples analysed. 

This free service is provided by the laboratory to all Clinitek 50 users.

Canterbury Health Laboratories recommends to all users of urine test strips that they regularly check their operators, test strips and procedure. 

For more information, please contact the POCT Coordinator.